An Article From:SCIENCE & MEDICINE
September/October 1996 · Volume 3, Number
5
Copyright © 1996 by Science & Medicine, Inc. All
Rights Reserved.
Focused Ultrasound Surgery Guided by MRI

By Kullervo Hynynen
Because it is sensitive to temperature changes in tissue, MRI is an effective
method for guiding and controlling ultrasound pulses. Ultrasound beams heat tissue, but
the objective of focused ultrasound therapy is tissue destruction, more like surgery than
hyperthermia, so it might be tumors in organs accessible to ultrasound or for coagulating
blood vessels. A prototype designed for breast tumor surgery is now being tested
clinically. Multiple ultrasound transducers in a phased array would permit focused
ultrasound surgery to be used for sizable tumors.
One of the dreams of medicine has been to be able to perform surgical procedures
without opening the patient. Non-invasive surgery would eliminate scar formation, blood
loss, and infections, and it would reduce the risk of other complications. Recovery time
would be shortened, and many procedures could be done on an outpatient basis. The absence
of tissue penetration would ease requirements for sterility in the operating room. The
benefits of non-invasive surgery would also result in a significant reduction in cost.
During the past 20 years this dream has come closer to reality because of the development
of imaging techniques that allow tumors and other anatomical structures to be visualized
with precision. The first successful application of non-invasive surgery was the use of
shock waves to disintegrate kidney stones. In this procedure, stones are located by x-ray
imaging and multiple shock waves are aimed at them through the skin and overlying tissues.
A stone is slowly broken into small fragments by the energy pulses, and the treatment is
continued until the x-rays verify that the stone has been destroyed.
This is an example of a completely noninvasive method that has revolutionized kidney stone
treatment and has resulted in the sorts of benefits already mentioned. Such as system
cannot be used to destroy soft tissues, but similar concepts have been explored for tumor
therapy. Recent advances have shown that such devices may be feasible and that noninvasive
surgery of soft tissues may become common practice within a few years.
One completely noninvasive method proposed for deep tissue destruction is focused
ultrasound. The sue of ultrasound beams as surgical tools was first proposed more than 50
years ago for destruction of brain tissue. A complex sonication system that used x-rays to
determine the target location with respect to the skull was developed by Bill and Frank
Fry at the University of Illinois during the 1950's and was tested in treatment of
Parkinson's disease, but it was never used outside the research setting.
The principle difficulty at the time was the accurate location of the target tissues. More
recently, focused ultrasound surgery systems have been combined with diagnostic ultrasound
imaging to make soft tissue tumor sonication possible. Several clinical trials for the
treatment of benign and malignant tumors of the prostate, bladder, kidney, and eye have
been conducted with these devices.
Although targeting with diagnostic ultrasound works well in many cases, the treatment
component still relies on open-loop energy delivery. Power settings are based on tests
done in animal tissues, and there is no on-line monitoring of the target location or the
magnitude of the temperature elevation. The treatment is therefore sensitive to variations
between patients, and the clinical results are variable.
 |
A temperature-sensitive
magnetic resonance image along the transducer axis shows focal temperature
elevation (arrow) induced by an ultrasound pulse in rabbit thigh muscle in vivo.
The scale is in centimeters. |
To eliminate these variations, energy delivery and its biological effects should be
monitored on line, and exposure should be adjusted to deliver comparable energy to all
patients. Researchers from the Brigham and Women's Hospital and Harvard Medical School
have worked with engineers and scientists from General Electric Medical Systems to develop
an ultrasound surgery system combined with magnetic resonance imaging. This makes on-line
temperature information available for monitoring and controlling energy delivery. The
status of this project is reviewed in the article.
Ultrasound Induces Thermal Effects
Ultrasound is a mechanical wave with a frequency above the audible range that propagates
by the motion of particles so that a pressure wave travels along with the mechanical
disturbance. The major advantage of using ultrasound for deep heating is that it
penetrates tissues well, at wavelengths on the order of a millimeter. An ultrasound beam
can be focused deep into the body onto a spot only a few millimeters in diameter. When the
beam is thus focused, ultrasound emitted by a transducer passes through the skin over a
wide area at intensities that cause no damage and then converges into a small spot at the
focus.
As an ultrasound save propagates through tissues, part of the energy is absorbed and
converted to thermal energy. The temperature elevation of the tissue caused by energy
absorption is inversely proportional to the beam area. The greatest temperature elevation
is induced at the focus, where it can be several hundred times more than in the overlying
tissues. This allows tissue at the focal spot to be selectively destroyed while
temperature elevation in surrounding tissues is mall, often less than 1°C.
Sharp focusing also allows fast energy delivery, so that temperature levels that cause
proteins to coagulate and cells to die can be reached in only a few seconds. The short
exposure produces sharp temperature gradients, and the transition distance between
coagulated cells and undamaged cells is only a few cells wide.
Ultrasound beams can be focused by using self-focusing radiators, lenses, or reflectors.
Transducers with many elements, each driven by properly delayed signals, so-called phased
arrays, can also be used to focus the beam. Electrical focusing using phased arrays
improves control over the ultrasound beam; for example, it can be used to simultaneously
generate multiple focal spots. All current clinical systems use spherically curved
transducers for energy delivery.
 |
Ultrasound beams may be focused
by curving the piezoelectric plate or by interposing a lens or reflector between a flat
plate and the target. A phased array of transducers is focused electronically. |
Several detailed studies have investigated the in vivo effects of ultrasound papameters
in animal tissue necrosis. Thermal or caviational effects that occur in tissues during
sonication were found to depend on the applied intensity, ultrasound frequency, exposure
duration, tissue type, and location. The temperature elevation of the tissue and the
duration of exposure determine the extent of tissue damage. Thresholds have been
established for many tissue and tumor types, showing that it is possible to kill any
tissue with a few seconds exposure if temperatures above 60 to 70°C are reached.
 |
Protein coagulation and
consequent tissue damage result from a combination of temperature elevation and
exposure duration. The graph shows the relationship between these factors. |
Cavitation is the formation and collapse of gas bubbles in tissue. Its mechanisms and
effects are not well understood and appear to depend on tissue type and location. Collapse
of the bubbles is associated with thigh temperatures and pressures, which result in
various degrees of mechanical damage to the tissue. Hemorrhage and blood vessel damage can
also occur when cavitation is present. In addition, the gas bubbles distort the
propagation of the ultrasound beam, causing uncertainty in determining the treated tissue
volume. The thermal mechanisms are much more predictable and better understood than
cavitation and have been used in most ultrasound surgery studies.
FOCUSED ULTRASOUND
Advantages
- Noninvasive
- Fast energy delivery:
| · |
Coagulated tissue volume is insensitive to
blood perfusion variations |
| · |
Tissues close to blood vessels can be
coagulated |
| · |
Coagulated tissue volume has sharp margins |
| · |
MRI monitoring of temperature elevation is not
strongly influenced by physiologic changes caused by the exposure |
- Good control over energy delivery results in accurate target
contouring and control over thermal exposure
Disadvantages
- Ultrasound beam is blocked by air or bone
- Patient to system coupling can be difficult
- Coagulation of large tissue volumes is slow with current
technology
|
INTERSTITIAL TECHNIQUES
(Laser, Radiofrequency, Hot Sources)
Advantages
- Simple energy delivery system
- Energy transmission to tissue through a small-diameter
probe:
| · |
Targeting is simple |
| · |
Organ motion is a smaller problem than with
external devices |
| · |
Treatment requires only a path for probe
access |
Disadvantages
- Slow energy delivery:
| · |
Perfusion-sensitive thermal exposures |
| · |
Margins of the treated zone are wider than
with ultrasound |
| · |
Blood vessels influence the shape of the
coagulated tissue volume |
| · |
Tissue close to blood vessels are difficult to
coagulate |
- Invasive; requires probe insertion
- Larger target volumes require multiple insertions
- Accurate coagulation of irregular tissue volumes is
difficult
- MRI thermometry is not accurate becuse long exposures allow
physiologic changes to affect the signal
|
Focused ultrasound surgery compares well with established minimally invasive techniques
such as lasers and radiofrequency ablation, which require a catheter to be inserted in the
tissue. With these systems, energy is delivered close to the catheter and spreads by
thermal conduction, so that adequate exposure takes several minutes. As a result, the
volume of coagulated tissue is variable and depends on the local blood flow and perfusion
of tissue.
Furthermore, because of wider thermal gradients, the transition from coagulated tissue to
normal tissue is less sharp with interstitial techniques than with ultrasound. Coagulation
of a large volume of tissue with interstitial techniques requires multiple needle
insertions. Irregular target volumes close to critical structures are also difficult to
treat with the same precision as can be achieved with focused ultrasound surgery.
In contrast, ultrasound is completely noninvasive and utilizes short exposures that result
in perfusion-insensitive temperature elevations with sharp gradients. Rapid energy input
makes tissue coagulation possible around large blood vessels, which would remove thermal
energy by convection during long exposures and thus protect tissues within a few
millimeters of the vessel walls The focal spot can be made small, and multiple sonications
allow good control over the volume of coagulated tissue, even close to critical
structures. Finally, the short ultrasound exposures allow temperature changes to be
detected quickly by MRI. The main disadvantage of focused ultrasound is that bone and gas
prevent beam propagation, so certain tissues such as lungs are difficult to treat.
 |
Change in the MR
signal during a 20 second test pulse in rabbit thigh in vivo. The phase change,
or proton resonance frequency shift (top panel), is a linear function of applied
power, even beyond the level of tissue necrosis, so tissue temperature can be reliably
estimated by this MR sequence. The signal intensity of the T1-weighted sequence (bottom
panel) is affected by tissue necrosis. |
 |
MRI Thermometry Localizes Focus and Controls Exposure
Magnetic resonance imaging is the only imaging method that provides adequate information
for guiding, monitoring, and controlling therapeutic interventions. The good tissue
contrast of MRI can be used for defining the target volume. Ultrasound surgery that can be
done while the scanner shows on-line images removes target tissue much more precisely than
is possible during open surgery. Equally important is the temperature sensitivity of some
MR sequences, allowing localization of the focus and control of the exposure. In addition,
tissue changes induced by temperature elevation are often visible in the MR images.
Noninvasive MRI thermometry makes use of the temperature dependence of some physical
property whose spatial distribution can be visualized. Three tissue properties have been
used for this purpose: spin-lattice decay time (T1), molecular diffusion of water
molecules, and the proton resonance frequency of water molecules. In vivo thermometry
based on T1 measurements during ultrasound therapy has proved difficult because of changes
such as edema and vasodilation induced in tissue by elevated temperatures. However, if
exposure times are in the 10 to 20 second range, these slow physiologic changes do not
have time to strongly influence tissue properties.
The diffusion coefficient technique that quantifies thermal Brownian movement has been
shown to work well in phantoms but is highly sensitive to tissue motion in vivo. Motion
sensitivity can be reduced by using echo planar imaging techniques that acquire images
quickly, but these require special hardware and reduce the signal to noise ratio.
Good temperature resolution has been obtained using the proton resonance frequency (PRF).
Changes in PRF induced by temperature elevation are linearly related to temperature and
can be mapped by using changes in phase images. The disadvantage of the frequency sift
technique is its insensitivity to temperature changes in fat.
The temperature dependence of T1-weighted and PRF (phase shift) images was tested in
rabbit thigh muscle in vivo as a method for monitoring noninvasive ultrasound surgery.
Both techniques could detect the temperature elevation, but the phase shift sequence
appeared to have a better ration of temperature contrast to noise. Both MR sequences are
sensitive enough to localize the temperature elevation and to monitor the normal tissue
exposure during therapy.
 |
Phase images across
the focal point in rabbit thigh muscle in vivo, illustrating the temperature elevation and
the effect of thermal conduction on the temperature during a 10-second sonication. The
four images (left to right were made approximately 2, 5.5, and 8.5 seconds from
the beginning of sonication and 5 seconds after the sound was turned off. |
Although in vivo tissues may have greater variation, the accuracy appears to be
adequate for localizing the focus with low-power test exposures, to assure that
temperatures between 60 and 100°C are reached during a 10-second therapeutic exposure,
and to monitor normal tissue temperature for safety. The upper temperature limit has been
set at 100°C to avoid boiling and the resulting formation of gas bubbles that could
distort the ultrasound beam. Thus the exposure limits are wide enough to accommodate some
degree of uncertainty in the temperature monitoring.
 |
T2-weighted images along the
axis of the ultrasound transducer after sonications of rabbit thigh muscle in
vivo illustrate the effect of sonicaton duration on coagulated tissue volume. Five
sonicatons at 56 watts and durations of 20, 10, 7, 5, and 3 seconds induced the lesions
shown from left to right. |
A Prototype System Is Being Tested
The feasibility of coagulating tissue under MRI guidance has been shown experimentally in
muscle, kidney, and brain tissue as well as in implanted tumors. In these tissues, the
test pulses are visible at power levels that do not cause tissue damage. In addition, the
tissue volume at the focus can be coagulated without damaging the overlying tissue. The
ability to control tissue damage in this fashion is especially important in the kidney,
which is highly perfused but lies beneath a layer of muscle that undergoes much less
convective cooling by blood flow.
Because propagation of an ultrasound beam is blocked by air or bone, brain sonications
require that a piece of bone be removed to create a window through which the beam can
pass.
 |
A postmortem photograph of a
rabbit kidney shows several coagulated tissue volumes after in vivo sonications. |
Besides destroying tissues, focused ultrasound can be used to occlude blood vessels.
The ultrasound exposures have not yet been optimized, but coagulation of capillaries and
larger arteries has been demonstrated in vivo.
Based on results with experimental system, a prototype ultrasound device for surgery of
breast tumors was manufactured by General Electric Medical Systems in collaboration with
members of the Department of Radiology at the Brigham and Women's Hospital and GE
Corporate Research & Development. The ultrasound fields are generated by a single,
focused, air-backed transducer mounted in a standard MRI table. The transducer can be
moved in the x, y, and z directions by a hydraulic positioning device within the water
bath that acts as a coupling medium. The workstation that controls the transducer is
programmed to aim the ultrasound beam at a location defined on an MR image.
 |
The focused ultrasound system
is enclosed in a plastic container filled with distilled degassed water, which serves as a
coupling medium. The container is covered with a plastic membrane and mounted within the
magnet in a standard MRI table. A flexible plastic bag filled with degassed water is
placed on top of the positioner under the patient's breast to improve acoustic coupling. A
surface coil enhances the signal to noise ratio. |
During a typical treatment, the target volume is first outlined on a series of MRI
scans. A low-energy test pulse is aimed at the center of the target volume by selecting
the location with a cursor. The workstation registers the target, aims the beam, generates
the test pulse, and transfers the temperature-sensitive image obtained during the
sonication. If the region of temperature elevation does not overlap the target volume, a
correction is made and a second test pulse is generated to verify the alignment accuracy.
After the test pulses, the complete target volume is sonicated using multiple pulses
placed so the coagulated volumes overlap. The dimensions of the coagulated tissue volume
for each sonication depend on the duration of the exposure and the applied power.
 |
Focused ultrasound treatment of
large tumors with MRI monitoring is a lengthy process. Many separately focused
sonications are required, each coagulating a small volume of tissue. A cooling period is
necessary after each exposure. A phased array of ultrasound transducers (bottom right)
could generate many focal spots simultaneously, in a pattern specifically tailored to the
tumor. |
Clinical testing of the prototype system in the treatment of the breast is now in
progress. MR images of tumors that had similar contrast uptake prior to treatment show
that treated tumors have no contrast uptake 48 hours later, whereas untreated tumors are
not affected. It is too early to draw conclusions about the clinical efficacy or toxicity
of the focused ultrasound treatment.
Focused ultrasound could replace some present surgical approaches to benign and malignant
tumors. Potential sites should have a soft tissue window that allows passage of the
ultrasound beam without encountering air or bone, and the breast is one of the best
examples.
Secondary liver cancer is a common problem with a poor prognosis, and liver tumors might
be amenable to focused ultrasound surgery. However, the available ultrasound window is
restricted by the ribs and by abdominal gas. Many prostate tumors could be reached through
the ultrasound window created by a full bladder, or rectal applicators could be developed.
Kidney tumors could also be treated, as could deep targets in the brain if a piece of
skull is removed.
Blood vessel occlusion is useful for treating arteriovenous malformations and some tumors
with an identifiable blood supply. It may also be helpful in the control of abdominal,
periotoneal, or pelvic hemorrhage and in the treatment of some trauma victims.
 |
A large target volume
in rabbit thigh muscle in vivo after multiple sonications covering a raster pattern. The
scale is in centimeters. |
Greater Focal Volume and Improved Temperature Monitoring Are Desirable
In all of the clinical ultrasound surgery studies to date, spherically curved, sharply
focused ultrasound transducers have been used. These effectively coagulate small volumes
of tissue, but the treatment of large tumors requires multiple sonications. Theoretical
and experimental studies have shown that a cooling period is required between exposures to
avoid temperature elevation in the tissue between the skin and the target. This makes the
treatment of large tumors long and expensive if MRI monitoring is used.
AN effective way to shorten the treatment time is to decrease the number of required
sonications. This can be achieved by increasing the acoustic focal volume. Ultrasound
phased arrays are the most flexible method, and they allow on line control of the
ultrasound beam. Electronic focusing has been used in diagnostic ultrasound for years.
Computer simulations have shown that arrays of hundreds of elements may be need when large
tumors are to be treated. These arrays cam generate many focal spots simultaneously and
tailor the sonication pattern for a given tumor, reducing treatment time and exposure to
surrounding tissues.
Phased arrays offer an additional benefit in that the focus can be changed electronically
instead of repositioning the transducer mechanically. This advantage is important in the
depth direction, where the magnet opening has limited space. Electronic focusing will also
be useful in the development of special applicators, such as intracavitary arrays for
surgery of the prostate.
 |
A rectangular volume of tissue
coagulated by multiple sonications has sharply defined boundaries, as seen in the
post-moretem phot of rabbit thign muscle. |
There is also room for improved MRI temperature monitoring. It is likely that once the
thermal surgery system is used routinely, additional resources will be dedicated to
improving the monitoring sequences. A significant improvement would be to obtain a
three-dimensional temperature map in the same time as the present 2D maps. A 3D map would
allow more precise on-line control to assure accurate target volume coverage and safety of
normal tissues. Fast on-line temperature calculation would allow the acoustic energy to be
controlled during each pulsed, so that accurate thermal exposures could be delivered
without repeated sonications.
Combining focused ultrasound beams with MRI allows precise and reliable destruction of
deep tissue volumes noninvasively. The technical feasibility of constructing such a
treatment system has been demonstrated, and the technique has shown promise in animal
experiments. However, clinical experiments are just beginning.
Treatment protocols have to be developed and tested in patients before wide clinical use
will be possible. Clinical trials will provide new information that will be used to
improve the equipment so that surgery can be safely and effectively performed. In
addition, the clinical experience will establish guidelines for the optimal use of this
new technology.
MRI-guided and monitored ultrasound surgery may eventually replace many open procedures by
offering increase precision while practically eliminating blood loss and disturbance to
normal tissue. This should translate to improved outcomes and significant cost reduction
because recover time and hospital stay will be shorter.
 |
Two breast tumors that
took up contrast material equally before focused ultrasound therapy show different
appearances in a T1-weighted contrast-enhanced image made seven days later. The treated
tumor (pink) shows no contrast uptake, but the untreated lesion closer to the
chest wall (green) shows enhancement. Lack of contrast enhancement correlates
with tissue coagulation. |
RECENT
REVIEWS
C.R. Hill and G.R. ter Haar: High intensity focused ultrasound -
potential for cancer treatment. British Journal of Radiology 68:1296-1301,
December 1995.
Ferenc A. Jolesz and S. Morry Blumenfeld: Interventional use of magnetic
resonance imaging. Magmetic Resonance Quarterly 10:85-96, 1995.
Narendra T. Sanghvi and Robert H. Hawes: High-intensity focused
ultrasound. Experimental and Investicational Endoscopy 4:383-395, April 1994.
ORIGINAL PAPERS
K. Hynynen et al., Medical Physics 20:107-115, January/
February 1993. [Feasibiligy study.] |
Harvey E. Cline et al., Radiology
194:732-737, March 1995. [Performance characteristics of the prototype.] Kagayaki Kuroda et al., Biomedical Thermology 13:43-62, 1995. [Water
proton magnetic resonance spectroscopic imaging.]
Yasutoshi Ishihara et al., Magnetic Resonance in Medicine
34:814-823, December 1995. [Temperature mapping using proton shift.]
Kullervo Hynynen et al., RadioGraphics 16:185-195, January
1996. [Details of the prototype system.]
Kullervo Hynynen et al., IEEE Transactions on Ultrasonics,
Ferroelectrics, and Frequency Control (accepted). [Feasibility of phased transducer
arrays.] |