advanced-surgical systems, inc
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 2 Channels with Independent Power and Frequency Control | |
| 50 Watts Maximum Power per Channel | |
| 1.5 - 2.0 MHz Frequency Range | |
| Pulsed Output intervals: 2-880ms | |
| Pulsed Output Duty-Cycle: 1-99% | |
| Forward and Reflected Power Monitoring | |
| Stand Alone Design (no external computer required) | |
| Unique Integrated Forward Power Feedback for Precise Power Delivery | |
| RS-232 Serial Interface and Real-Time Parallel Control Lines | |
| Watchdog Timers to Prevent Unwanted Power Delivery | |
| RF Isolation Transformers | |
| Simple "English Language" Programming Interface | |
| Microsoft Windows® Sample Interface Including Source-Code | |
| One Year Full-Parts and Labor Warranty | |
| Unlimited No-Charge Technical Support |
The 500-023 is a specialized version of the 500-009 CSA.
The primary system differences are listed in the following table:
| Parameter | 500-023 Pulsed Output | 500-009 CSA |
| Output Power | 50W | 50W |
| Frequency Range: | 1500-2000 kHz | 6000-10000 kHz |
| Channels | 2 | 4 |
| Power Set Resolution | 8-bits | 8-bits |
| Power Measurement Resolution | 8-bits | 8-bits |
| Frequency Modulation | No | Yes |
| Pulsed Output | Yes | No |
| Pulsed Output Interval | 2 - 8 ms | N/A |
| Pulsed Output Duty-Cycle | 1 - 99% | N/A |
The 500-023 is an ultrasound driving system (UDS) designed to generate an RF signals necessary to drive ultrasound transducers in the 1.5 - 2.0 MHz frequency range, supplying up to 50W per channel of RF power. The 500-023 is a self-contained four-channel driving system that allows complete control of the amplifiers via a front panel keypad, or via an RS-232 computer interface. All channels of the system may be set up and operated independently. In addition, under software control, all channels may be enabled or disabled simultaneously. Each channel retains its own settings, and the outputs are all gated on or off together. When Pulsing is enabled the amplifier will switch on and off in accordance with the pulse interval and duty cycle settings.
| Parameter | Value |
|
Amplifier |
|
|
Number of Channels |
2 |
|
Output Power per Channel |
50 Watts Max. |
|
Power Control Resolution |
8 bits |
|
Power Control Accuracy |
±1% |
|
Power Measurement |
|
|
Resolution |
8 bits |
|
Accuracy |
±1% |
|
Frequency Synthesizer |
|
|
Range |
1.5 - 2.0 MHz |
|
Resolution |
~0.01 MHz |
|
Accuracy |
±0.8% at 7 MHz |
|
Stability |
±100 ppm |
The 500-023 will be warranted for a period of one year, including all parts and labor, providing the system is operated within the specified operating range. Additionally, Advanced Surgical Systems is willing to work with its customers to make sure that the equipment is used to its maximum potential, whether this be regarding interfacing with the transducers, developing control software, or any other issues that may arise throughout the lifetime of the 500-023 system. Throughout the life of the product, technical support is available for no charge.
The systems as quoted consist only of the systems described herein. Any modifications may affect the actual pricing. Please be advised that the pricing information shown here is not to be considered a formal quote. For a formal quote, please contact Advanced Surgical Systems, Inc.
| Description | SKU # | Price |
|---|---|---|
| 500-023 | 500-023 | $9995 |
| Delivery: | 4 mo. ARO |
| Warranty: | 1 year, full parts and labor |
| F.O.B.: | Advanced Surgical Systems, Inc., Tucson, AZ. |
| Payment Terms: | 50% Down with Purchase Order, Remainder Due on Delivery |